Network and Remote Operation
R&S
®
FSW
776
User Manual 1173.9411.02 ─ 43
Recording is stopped when you deactivate "Auto Recording".
To continue recording, reactivate "Auto Recording".
To start a new SCPI command list, select
before activating "Auto Record-
ing".
Note:
●
Some parameters cannot be set by a SCPI command.
In this case,
no SCPI command found
is entered in the list instead of a command
when you record settings automatically.
●
The R&S
FSW automatically clears the SCPI command list after booting.
List of recorded commands / script editor
The currently recorded commands are displayed in a basic editor directly in the SCPI
Recorder dialog box. Right-click the editor to display a context-sensitive menu with
basic editing functions for the list, such as copy, paste, delete, undo and redo.
Remove Last
Deletes the last recorded SCPI command from the list.
Load Recording
Loads an existing script in ASCII format (
*.inp
) from a file to the script editor. If the
editor contains recorded commands, you must confirm a message to overwrite them. A
file selection dialog box is displayed.
Save As
Saves the current SCPI command list to the
C:\R_S\INSTR\ScpiRecordings
directory in ASCII format with the file extension
.inp
.
Tip:
You can execute the command list in an
.inp
file without further editing using the
IECWIN tool provided with the R&S
FSW, see
Chapter 13.3, "The IECWIN Tool"
on page 771. You can also reload
.inp
files to the script editor later.
Export
Exports the current SCPI command list to the specified file and directory in the
selected format. By default, the file is stored in the
C:\R_S\INSTR\ScpiRecordings
directory. Besides the recorded commands them-
selves, the exported script includes all format-specific header data required to execute
the script using an external program on the controller.
Automating Tasks with Remote Command Scripts