Remote Commands
R&S
®
FSW
1287
User Manual 1173.9411.02 ─ 43
For details, see
Chapter 12.4, "Frequency Response Correction (R&S
Suffix:
<si>
.
1..n
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Example:
SENS:CORR:FRES:USER:STAT
Manual operation:
See
[SENSe:]CORRection:FRESponse<si>:BASeband:USER:STORe
<FilePath>
[SENSe:]CORRection:FRESponse<si>:INPut<ip>:USER:STORe
<FilePath>
[SENSe:]CORRection:FRESponse<si>:USER:STORe
<FilePath>
Stores a saveset for a user-defined frequency response correction scenario. By
default, the saveset is stored in the
C:\R_S\INSTR\USER\FResponse
directory.
Suffix:
<si>
.
1..n
irrelevant
Setting parameters:
<FilePath>
string
Example:
SENS:CORR:FRES:USER:STOR 'FRes1'
Stores the current frequency response correction configuration
to a file named
C:\R_S\INSTR\USER\FResponse\FRes1.dfl
.
Usage:
Setting only
Manual operation:
See
[SENSe:]CORRection:FRESponse<si>:USER:VALid?
This command queries the validity of the user-defined correction settings.
Suffix:
<si>
.
1..n
irrelevant
Return values:
<Validity>
0 | 1
1
The setting is valid.
Configuring the R&S
FSW