Remote Commands
R&S
®
FSW
1255
User Manual 1173.9411.02 ─ 43
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see
Chapter 5.1.7, "Protecting Data Using the Secure User Mode"
Suffix:
<n>
.
Parameters:
<FileName>
String containing the path and name of the target file.
Example:
MMEM:STOR:SGR 'Spectrogram'
Copies the spectrogram data to a file.
Manual operation:
See
MMEMory:STORe<n>:SPURious
<FileName>
This command exports the marker peak list available for spurious emission measure-
ments to a file.
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see
Chapter 5.1.7, "Protecting Data Using the Secure User Mode"
Suffix:
<n>
.
irrelevant
Parameters:
<FileName>
String containing the path and name of the target file.
Example:
MMEM:STOR:SPUR 'test'
Saves the current marker peak list in the file test.dat.
14.9.6
Examples: Managing Data
.................................................................................1257
............................................................................... 1257
Storing Multiple Graphical Measurement Results to a PDF File
Managing Settings and Results