Remote Commands
R&S
®
FSW
1116
User Manual 1173.9411.02 ─ 43
[SENSe:]PMETer<p>:TRIGger:LEVel
..............................................................................1117
[SENSe:]PMETer<p>:TRIGger:SLOPe
............................................................................1117
[SENSe:]PMETer<p>:TRIGger[:STATe]
........................................................................... 1117
[SENSe:]PMETer<p>:TRIGger:DTIMe
<Time>
This command defines the time period that the input signal has to stay below the IF
power trigger level before the measurement starts.
Suffix:
<p>
.
Power sensor index
Parameters:
<Time>
Range:
0 s to 1 s
Increment: 100 ns
*RST:
100 µs
Default unit: S
Example:
PMET2:TRIG:DTIMe 0.001
[SENSe:]PMETer<p>:TRIGger:HOLDoff
<Holdoff>
This command defines the trigger holdoff for external power triggers.
Suffix:
<p>
.
Power sensor index
Parameters:
<Holdoff>
Time period that has to pass between the trigger event and the
start of the measurement, in case another trigger event occurs.
Range:
0 s to 1 s
Increment: 100 ns
*RST:
0 s
Default unit: S
Example:
PMET2:TRIG:HOLD 0.1
Sets the holdoff time of the trigger to 100 ms
Manual operation:
See
[SENSe:]PMETer<p>:TRIGger:HYSTeresis
<Hysteresis>
This command defines the trigger hysteresis for external power triggers.
The hysteresis in dB is the value the input signal must stay below the IF power trigger
level in order to allow a trigger to start the measurement.
Suffix:
<p>
.
Power sensor index
Parameters:
<Hysteresis>
Range:
3 dB to 50 dB
Increment: 1 dB
*RST:
0 dB
Default unit: DB
Setting Basic Measurement Parameters