Remote Commands
R&S
®
FSW
1096
User Manual 1173.9411.02 ─ 43
Manual operation:
See
SOURce<si>:EXTernal<gen>:FREQuency:COUPling[:STATe]
<State>
This command couples the frequency of the external generator output to the
R&S
FSW.
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<State>
ON | OFF | 0 | 1
ON | 1
Default setting: a series of frequencies is defined (one for each
sweep point), based on the current frequency at the RF input of
the R&S
FSW; the RF frequency range covers the currently
defined span of the R&S
FSW (unless limited by the range of the
signal generator)
OFF | 0
The generator uses a single fixed frequency, defined by
SOURce<si>:EXTernal<gen>:FREQuency
*RST:
1
Example:
SOUR:EXT:FREQ:COUP ON
Manual operation:
See
SOURce<si>:EXTernal<gen>:FREQuency[:FACTor]:DENominator
<Value>
This command defines the denominator of the factor with which the analyzer frequency
is multiplied in order to obtain the transmit frequency of the selected generator.
Select the multiplication factor such that the frequency range of the generator is not
exceeded if the following formula is applied to the start and stop frequency of the ana-
lyzer:
Offset
Analyzer
Generator
F
ator
Deno
Numerator
F
F
min
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<Value>
<numeric value>
*RST:
1
Setting Basic Measurement Parameters