Remote Commands
R&S
®
FSW
1035
User Manual 1173.9411.02 ─ 43
[SENSe:]SWEep:DURation?
<Time>
This command provides an estimation of the total time required to capture the data and
process it. This time span may be considerably longer than the actual sweep time (see
Tip:
To determine the necessary timeout for data capturing in a remote control pro-
gram, double the estimated time and add 1 second.
Return values:
<Time>
Example:
SWE:TIME 1s
SWE:DUR?
Reply:
27.9734842578
Usage:
Query only
Manual operation:
See
See
"Data capturing takes too long"
[SENSe:]SWEep:FFTSubspan?
Returns the number of FFT partial spans required to cover the entire measurement
range (read-only). See also
Only available in FFT sweep mode in the Spectrum application, and not for SEM,
ACLR, or Spurious emissions measurements.
Return values:
<NoOfPartialSpans> integer
Usage:
Query only
Manual operation:
See
[SENSe:]SWEep:OPTimize
<Mode>
In FFT mode, several FFT analysis steps are required to cover the entire measurement
span. The span which is covered by one FFT analysis step is called
subspan
. The sub-
span cannot be defined directly, but it can be optimized according to measurement
requirements.
Setting Basic Measurement Parameters