
Remote Commands
R&S
®
FSW
1012
User Manual 1173.9411.02 ─ 43
<MeasTime>
Defines the measurement time.
Range:
1 µs to 30 s
Default unit: S
<TriggerSource>
EXTernal | EXT2 | EXT3 | VIDeo
Selects a trigger source.
For more information see
Configuring Triggered and Gated Mea-
<TriggerLevel>
Defines a trigger level.
The trigger level is available for the video trigger. In that case,
the level is a percentage of the diagram height.
In case of an external trigger, the R&S
FSW uses a fix TTL level.
Range:
0 to 100
Default unit: PCT
<TriggerOffset>
Defines the trigger delay.
Range:
0 s to 30 s
*RST:
0 s
Default unit: s
<Detector>
Selects the detector and therefore the way the measurement is
evaluated.
MEAN
Calculates the RMS pulse power.
PEAK
Calculates the peak pulse power.
<OfPulses>
Defines the number of pulses included in the measurement.
Range:
1 to 32001
Return values:
<PowerLevel>
Pulse power level.
The result is a list of floating point values separated by commas.
The unit is dBm.
[SENSe:]MPOWer:RESult:MIN?
This command queries the lowest pulse power that has been measured during a pulse
power measurement.
If there are no results, the command returns an error.
Return values:
<PulsePower>
Lowest power level of the pulse power measurement.
The unit is dBm.
Usage:
Query only
14.5.16.2
Example: Performing a Pulse Power Measurement
The following example shows a pulse power measurement.
Configuring and Performing Measurements