Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
311
User Manual 1173.9240.02 ─ 23
Parameters:
<Separator>
COMMa
Uses a comma as decimal separator, e.g.
4,05
.
POINt
Uses a point as decimal separator, e.g.
4.05
.
*RST:
*RST has no effect on the decimal separator.
Default is POINt.
Example:
FORM:DEXP:DSEP POIN
Sets the decimal point as separator.
Manual operation:
See
See
FORMat:DEXPort:HEADer
<State>
If enabled, additional instrument and measurement settings are included in the header
of the export file for result data. If disabled, only the pure result data from the selected
traces and tables is exported.
See
Chapter 11.7.5, "Reference: ASCII File Export Format"
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Manual operation:
See
"Include Instrument Measurement Settings"
FORMat:DEXPort:TRACes
<Selection>
This command selects the data to be included in a data export file (see
Parameters:
<Selection>
SINGle
Only a single trace is selected for export, namely the one speci-
fied by the
command.
ALL
Selects all active traces and result tables (e.g. Result Summary,
marker peak list etc.) in the current application for export to an
ASCII file.
The <trace> parameter for the
command is ignored.
*RST:
SINGle
Manual operation:
See
"Export all Traces and all Table Results"
Retrieving Results