![R&S FSW-K7 User Manual Download Page 286](http://html1.mh-extra.com/html/rands/fsw-k7/fsw-k7_user-manual_3082493286.webp)
Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
286
User Manual 1173.9240.02 ─ 23
Parameters:
<State>
ON | 1
The measurement for automatic adjustment waits for the trigger.
OFF | 0
The measurement for automatic adjustment is performed imme-
diately, without waiting for a trigger.
*RST:
1
[SENSe:]ADJust:FREQuency
This command sets the center frequency to the frequency with the highest signal level
in the current frequency range.
Example:
ADJ:FREQ
Usage:
Event
Manual operation:
See
"Adjusting the Center Frequency Automatically (Auto Freq)"
[SENSe:]ADJust:LEVel
This command initiates a single (internal) measurement that evaluates and sets the
ideal reference level for the current input data and measurement settings. This ensures
that the settings of the RF attenuation and the reference level are optimally adjusted to
the signal level without overloading the R&S
FSW or limiting the dynamic range by an
S/N ratio that is too small.
Example:
ADJ:LEV
Usage:
Event
Manual operation:
See
"Setting the Reference Level Automatically (Auto Level)"
[SENSe:]ADJust:SCALe[:Y]:AUTO[:CONTinuous]
<State>
Activates automatic scaling of the y-axis in all diagrams according to the current mea-
surement results. Currently auto-scaling is only available for AF measurements. RF
power and RF spectrum measurements are not affected by the auto-scaling.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
SENS:ADJ:SCAL:Y:AUTO ON
Manual operation:
See
Configuring the Measurement