Parameters:
<Attenuation>
attenuation in dB
Range:
see data sheet
Increment: 1 dB
*RST:
0 dB (OFF)
Example:
INP:EATT:AUTO OFF
INP:EATT 10 dB
Manual operation:
See
"Using Electronic Attenuation"
INPut:EATT:AUTO
<State>
This command turns automatic selection of the electronic attenuation on and off.
If on, electronic attenuation reduces the mechanical attenuation whenever possible.
This command requires the electronic attenuation hardware option.
It is not available if the optional Digital Baseband Interface is active.
Parameters:
<State>
1 | 0 | ON | OFF
1 | ON
0 | OFF
*RST:
1
Example:
INP:EATT:AUTO OFF
Manual operation:
See
"Using Electronic Attenuation"
INPut:EATT:STATe
<State>
This command turns the electronic attenuator on and off.
This command requires the electronic attenuation hardware option.
It is not available if the optional Digital Baseband Interface is active.
Parameters:
<State>
1 | 0 | ON | OFF
1 | ON
0 | OFF
*RST:
0
Example:
INP:EATT:STAT ON
Switches the electronic attenuator into the signal path.
Manual operation:
See
"Using Electronic Attenuation"
Configuring the Measurement