Analysis
R&S
®
FSW-K7
136
User Manual 1173.9240.02 ─ 23
Remote command:
Average Count
Determines the number of averaging or maximum search procedures If the trace
modes "Average", "Max Hold" or "Min Hold" are set.
In continuous sweep mode, if sweep count = 0 (default), averaging is performed over
10 sweeps. For sweep count =1, no averaging, Max Hold or Min Hold operations are
performed.
Remote command:
Predefined Trace Settings - Quick Config
Commonly required trace settings have been predefined and can be applied very
quickly by selecting the appropriate button.
Function
Trace Settings
Preset All Traces
Trace 1:
Clear Write
Traces 2-6:
Blank
Set Trace Mode
Max | Avg | Min
Trace 1:
Max Hold
Trace 2:
Average
Trace 3:
Min Hold
Traces 4-6:
Blank
Set Trace Mode
Max | ClrWrite | Min
Trace 1:
Max Hold
Trace 2:
Clear Write
Trace 3:
Min Hold
Traces 4-6:
Blank
Trace 1/Trace 2/Trace 3/Trace 4 (Softkeys)
Displays the "Traces" settings and focuses the "Mode" list for the selected trace.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>[:STATe]
6.2 Trace / Data Export Configuration
Access
: "Save" > "Export" > "(Trace) Export Config"
Or
: TRACE > "Trace Config" > "Trace/Data Export"
The standard data management functions (e.g. saving or loading instrument settings)
that are available for all R&S
FSW applications are not described here.
See the R&S
FSW User Manual for a description of the standard functions.
Trace / Data Export Configuration