Network and Remote Operation
R&S
®
FSVA3000/ R&S
®
FSV3000
729
User Manual 1178.8520.02 ─ 01
The changes in the device settings and the recorded measurement values are dis-
played on the instrument screen.
3. To obtain optimum performance during remote control, send the
SYSTem:DISPlay:UPDate OFF
command to hide the display of results and dia-
grams again (default setting in remote control).
4. To prevent unintentional return to manual operation, disable the keys of the instru-
ment using the universal command
LLO
.
Switching to manual mode is only possible via remote control then. This function is
only available for the GPIB interface.
5. To enable the keys of the R&S
FSV/A again, switch the instrument to local mode
(
GTL
- Go to Local), i.e. deactivate the
REN
line of the remote control interface.
If the instrument is operated exclusively in remote control, it is recommended that you
switch off the display. For details see
13.6.9
How to Return to Manual Operation
Before you switch back to manual operation, all remote command processing must be
completed. Otherwise, the instrument will switch back to remote control immediately.
► Select the "Local" softkey, or use the following GPIB command:
status = viGpibControlREN(vi, VI_GPIB_REN_ADDRESS_GTL)
If you select the "Local" softkey while a self-alignment or a self-test is still running
(which was started remotely), the instrument only returns to the manual operation state
when the alignment or test is completed.
How to Set Up a Network and Remote Control