Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
199
User Manual 1178.8536.02 ─ 01
This value can only be determined in triggered measurements using external or
IFPower triggers, otherwise the value is 0.
Return values:
<TPIS>
numeric value
Default unit: s
Example:
TRAC:IQ:TPIS?
Result for a sample rate of 1 MHz: between 0 and 1/1 MHz, i.e.
between 0 and 1
μ
s (the duration of 1 sample).
Usage:
Query only
Manual operation:
See
TRACe:IQ:WBANd[:STATe]
<State>
This command determines whether the wideband provided by bandwidth extension
options is used or not (if installed).
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation:
See
TRACe:IQ:WBANd:MBWidth
<Limit>
Defines the maximum analysis bandwidth. Any value can be specified; the next higher
fixed bandwidth is used.
Defining a value other than "MAX" is useful if you want to specify the sample rate
directly and at the same time, ensure a minimum bandwidth is available.
(See
Chapter 4.1.1, "Sample Rate and Maximum Usable I/Q Bandwidth for RF Input"
Parameters:
<Limit>
40 MHz
Restricts the analysis bandwidth to a maximum of 40
MHz. The
bandwidth extension option R&S
FSV3-B200 or B400 is deacti-
vated.
200 MHz
Restricts the analysis bandwidth to a maximum of 200
MHz. The
bandwidth extension option R&S
FSV3-B400 is deactivated.
*RST:
maximum available
Default unit: Hz
Configuring I/Q Analyzer Measurements