OUTPut<up>:IF:IFFRequency
<Frequency>
This command defines the frequency for the IF output of the R&S
FSV/A. The IF fre-
quency of the signal is converted accordingly.
This command is available in the time domain and if the IF output is configured for IF.
Suffix:
<up>
.
Parameters:
<Frequency>
*RST:
50.0 MHz
Default unit: HZ
OUTPut<up>:IF:STATe
<State>
Enables or disables output of the measured IF value at the "IF" output connector.
Suffix:
<up>
.
1..n
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Example:
OUTP:IF:STAT ON
OUTPut<up>:UPORt:STATe
<State>
This command toggles the control lines of the user ports for the
AUX PORT
connector.
This 9-pole SUB-D male connector is located on the rear panel of the R&S
FSV/A.
For the R&S
FSV/A, this command is maintained for compatibility reasons only. The
AUX PORT
connector only supports output.
Suffix:
<up>
.
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
User port is switched to INPut
ON | 1
User port is switched to OUTPut
Example:
OUTP:UPOR:STAT ON