Analysis
R&S
®
FSV/A3000 I/Q
Analyzer
123
User Manual 1178.8536.02 ─ 01
For each listed peak the frequency/time ( "X-value" ) and level ( "Y-Value" ) values are
given.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:STATe
Sort Mode
Defines whether the peak list is sorted according to the x-values or y-values. In either
case the values are sorted in ascending order.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:SORT
Maximum Number of Peaks
Defines the maximum number of peaks to be determined and displayed.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:LIST:SIZE
Peak Excursion
Defines the minimum level value by which a signal must rise or fall so that it is identi-
fied as a maximum or a minimum by the search functions.
Entries from 0
dB to 80
dB are allowed; the resolution is 0.1
dB. The default setting for
the peak excursion is 6
dB.
For analog demodulation, the unit and value range depend on the selected result dis-
play type.
Remote command:
CALCulate<n>:MARKer<m>:PEXCursion
Display Marker Numbers
By default, the marker numbers are indicated in the diagram so you can find the peaks
from the list. However, for large numbers of peaks the marker numbers may decrease
readability; in this case, deactivate the marker number display.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:ANNotation:LABel[:STATe]
Export Peak List
The peak list can be exported to an ASCII file (
.DAT
) for analysis in an external appli-
cation.
Remote command:
6.4.4.3
Deactivating All Marker Functions
Access
: "Overview" > "Analysis" > "Marker Functions" > "All Functions Off"
All special marker functions can be deactivated in one step.
Marker Usage