Instrument Functions for Bluetooth Measurements (K8)
R&S
®
FSV-K8
62
Operating Manual 1176.7610.02 ─ 02
●
●
●
Note:
For R&S power sensors, the "Gate Mode"
Lvl
is not supported. The signal sent
by these sensors merely reflects the instant the level is first exceeded, rather than a
time period. However, only time periods can be used for gating in level mode. Thus,
the trigger impulse from the sensors is not long enough for a fully gated measurement;
the measurement cannot be completed.
Remote command:
TRIG:SOUR PSE
, see
Time ← Trg/Gate Source
Opens an edit dialog box to define a repetition interval in which the measurement is
triggered. The shortest interval is 2 ms.
Remote command:
TRIG:SOUR TIME
Digital IQ ← Trg/Gate Source
For I/Q Analyzer or AnalogDemod mode only:
Defines triggering of the measurement directly via the LVDS connector. In the sub-
menu you must specify which general purpose bit (GP0 to GP5) will provide the trigger
data.
This trigger mode is available for input from the R&S Digital I/Q Interface (option
R&S
FSV-B17) only.
The following table describes the assignment of the general purpose bits to the LVDS
connector pins.
Table 3-1: Assignment of general purpose bits to LVDS connector pins
Bit
LVDS pin
GP0
SDATA4_P - Trigger1
GP1
SDATA4_P - Trigger2
GP2
SDATA0_P - Reserve1
GP3
SDATA4_P - Reserve2
GP4
SDATA0_P - Marker1
GP5
SDATA4_P - Marker2
Remote command:
TRIG:SOUR GP0
, see
Trg/Gate Level
Opens an edit dialog box to enter the trigger/gate level.
In the trigger modes "Time" and "Power Sensor", this softkey is not available.
Remote command:
Softkeys of the Trigger Menu (Option R&S
FSV-K8)