Remote Commands for Bluetooth Measurements (K8)
R&S
®
FSV-K8
123
Operating Manual 1176.7610.02 ─ 02
Example:
DDEM:SEAR:TIME 100US
Sets the record length for sync word and burst search to 100 us.
Mode:
BT
Manual operation:
See
Table 4-4: Max. record length depending on points per symbol
Points per Symbol
Maximum record length
2
104.4 slots
4
52.2 slots
8
26.1 slots
16
13.1 slots
32
6.5 slots
[SENSe:]DDEMod:SEARch:TIME:AUTO
<Mode>
This command activates the automatic setting of the record length for the sync word
and burst search, depending on the selected packet type.
The automatic record length is determined a follows:
Trigger free run:
search length = 3 * packet | sync offset or burst offset |
All other trigger modes:
search length = 1 * packet 1 Slot + | sync offset or burst offset |
If the selected measurement time is higher than the packet length, the following differ-
ence is added to the search length:
measurement time – packet length
Parameters:
<Mode>
ON | OFF
*RST:
ON
Example:
DDEM:SEAR:TIME:AUTO OFF
Selects manual input for the record length
Mode:
BT
Manual operation:
See
4.11 SYSTem subsystem
SYSTem:DISPlay:UPDate
<State>
In remote control mode, this command switches on or off the instrument display. If
switched on, only the diagrams, traces and display fields are displayed and updated.
SYSTem subsystem