FM Stereo Option R&S
FSV–K7S
R&S
®
FSV-K7/K7S
225
Operating Manual 1176.7561.02 ─ 06
CALC:MARK1:FUNC:NDBD:FREQ?
(span > 0), see
CALC:MARK1:FUNC:NDBD:TIME?
(span = 0), see
Marker Peak List
Opens the "Peak List" submenu to define criteria for the sort order and the contents of
the peak list. For each listed peak the frequency ("Stimulus") and level ("Response")
values are given. In addition, the peaks are indicated in the trace display. A maximum
of 50 entries are listed.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:COUNt?
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:X
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:Y?
Peak List On/Off ← Marker Peak List
Activates/deactivates the marker peak list. If activated, the peak list is displayed and
the peaks are indicated in the trace display.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:STAT
Sort Mode Freq/Lvl ← Marker Peak List
Defines the criteria for sorting:
"Freq"
sorting in ascending order of frequency values (span > 0) or time val-
ues (span = 0)
"Lvl"
sorting in ascending order of the level
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:SORT
Max Peak Count ← Marker Peak List
Defines the maximum number of peaks to be determined and displayed.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:LIST:SIZE
Peak Excursion ← Marker Peak List
Opens an edit dialog box for level measurements to enter the minimum level value by
which a signal must rise or fall so that it will be identified as a maximum or a minimum
by the search functions. Entries from 0 dB to 80 dB are allowed; the resolution is 0.1
dB. The default setting for the peak excursion is 6 dB.
For more information see "Specifying the suitable peak excursion" and "Effect of differ-
ent peak excursion settings".
Remote command:
CALCulate<n>:MARKer<m>:PEXCursion
Instrument Functions FM Stereo (R&S
FSV–K7S)