Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
164
Operating Manual 1176.7561.02 ─ 06
TRIGger<n>[:SEQuence]:IFPower:HYSTeresis
<Value>
This command sets the limit that the hysteresis value for the IF power trigger has to fall
below in order to trigger the next measurement.
Suffix:
<n>
.
irrelevant
Parameters:
<Value>
*RST:
3 dB
Example:
TRIG:SOUR IFP
Sets the IF power trigger source.
TRIG:IFP:HYST 10DB
Sets the hysteresis limit value.
Manual operation:
See
TRIGger<n>[:SEQuence]:LEVel:AM[:ABSolute]
<Level>
The command sets the level when RF power signals are used as trigger source.
For triggering with AF, AM, AMRelative, FM, and PM trigger sources to be successful,
the measurement time must cover at least 5 periods of the audio signal.
Suffix:
n
.
irrelevant
Parameters:
<Level>
Range:
-100 to +30
*RST:
-20 dBm
Default unit: dBm
Example:
TRIG:LEV:AM -30 dBm
Sets the RF power signal trigger threshold to -30 dBm
Mode:
ADEMOD
Manual operation:
See
TRIGger<n>[:SEQuence]:LEVel:AM:RELative
<Level>
The command sets the level when AM-modulated signals are used as trigger source.
For triggering with AF, AM, AMRelative, FM, and PM trigger sources to be successful,
the measurement time must cover at least 5 periods of the audio signal.
Suffix:
n
.
irrelevant
Parameters:
<Level>
Range:
-100 to +100
*RST:
0 %
Default unit: %
Remote Commands of the Analog Demodulation (R&S
FSV–K7)