Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
154
Operating Manual 1176.7561.02 ─ 06
Parameters:
<Frequency>
Range:
0 to fmax
*RST:
fmax/2
Default unit: Hz
f
max
is specified in the data sheet. min span is 10 Hz
Example:
FREQ:CENT 100 MHz
Manual operation:
See
[SENSe:]FREQuency:CENTer:STEP
<StepSize>
This command defines the center frequency step size.
Parameters:
<StepSize>
Range:
1 to fmax
*RST:
0.1 x <span value>
Default unit: Hz
Example:
FREQ:CENT:STEP 120 MHz
Manual operation:
See
[SENSe:]FREQuency:CENTer:STEP:LINK
<CouplingType>
This command couples and decouples the center frequency step size to the span or
the resolution bandwidth.
Parameters:
<CouplingType>
SPAN
Couples the step size to the span. Available for measurements
in the frequency domain.
(for RF spectrum result display)
RBW
Couples the step size to the resolution bandwidth. Available for
measurements in the time domain.
(for all result displays except RF spectrum)
OFF
Decouples the step size (manual input).
*RST:
SPAN
Example:
FREQ:CENT:STEP:LINK SPAN
Manual operation:
See
See
"0.1*Demod BW (AF/RF Time Domain, AF Spectrum)"
on page 41
See
See
"0.5*Demod BW (AF/RF Time Domain, AF Spectrum)"
on page 41
See
See
"x*Demod BW (AF/RF Time Domain, AF Spectrum)"
Remote Commands of the Analog Demodulation (R&S