Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
152
Operating Manual 1176.7561.02 ─ 06
[SENSe:]FILTer<n>:HPASs:FREQuency
<FilterType>
This command selects the high pass filter type in the specified window. For details on
filters refer to
For details about the demodulation bandwidth range refer to
Suffix:
<n>
.
1...4
window
Parameters:
<FilterType>
Range:
50 to 300
*RST:
300Hz
Default unit: Hz
Example:
FILT:HPAS:FREQ 300Hz
Selects the high pass filter for the demodulation bandwidth
range from 800 Hz to 16 MHz.
Mode:
ADEMOD
Manual operation:
See
[SENSe:]FILTer<n>:LPASs[:STATe]
<State>
This command activates/deactivates the selected low pass filter in the specified win-
dow.
For details on the low pass filter refer to
Suffix:
<n>
.
1...4
window
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
FILT:LPAS ON
Activates the selected low pass filter.
Mode:
ADEMOD
Manual operation:
See
[SENSe:]FILTer<n>:LPASs:FREQuency[:ABSolute]
<FilterType>
This command selects the absolute low pass filter type in the specified window. For
details on filters refer to
For details about the demodulation bandwidth range refer to
Suffix:
<n>
.
1...4
window
Remote Commands of the Analog Demodulation (R&S