Measurements and Result Displays
R&S
®
FSPN
90
User Manual 1179.4363.02 ─ 01
Result
Description
Trace
Shows the number of the trace the integration is
applied to.
Start / Stop Offset
Shows the start and stop offset of the integration
interval.
Weighting
Shows the name of the weighting filter, if you have
applied one.
Int Noise
Shows the integrated noise.
The integral is calculated over the frequency range
defined by the "Start" and "Stop" values.
PM
*
Shows the residual PM result in degrees and rad.
traces.)
FM / AM
*
Shows the residual FM results in Hz or the residual
AM results in %, depending on the
traces.)
Jitter
*
Shows the jitter in seconds.
traces.)
* For baseband noise measurements, these values are always 0.
Remote command:
Int. PHN:
FETCh<n>[:RANGe<j>]:PNOise<t>:IPN?
FETCh<n>[:RANGe<j>]:PNOise<t>:RFM?
AM:
FETCh<n>[:RANGe<j>]:PNOise<t>:RAM?
PM:
FETCh<n>[:RANGe<j>]:PNOise<t>:RPM?
Jitter:
FETCh<n>[:RANGe<j>]:PNOise<t>:RMS?
Spurious List
are peak levels at one or more offset frequencies and are caused mostly by
interfering signals. The "Spurious List" result display shows the location of all detected
spurs in a table.
Note that only signals above a certain
are regarded as spurs. This threshold
is also considered in the spurious list if spur removal has been turned off for a trace.
The order of the spurs in the table depends on the
.
The table consists of a variable number of rows. For each detected spur, the table
shows several results.
Wnd
Shows the number of the measurement window the
spur is in (usually "1", unless you have several noise
diagrams open at the same time).
Trace
Shows the trace that the spur is on.
Spur
Shows the spur number. Spurs are sorted by their
frequency, beginning with the spur with the lowest
frequency.
Offset
*
Shows the position (offset frequency) of the spur.
Result Displays