Remote Commands
R&S
®
FSMR3000
458
User Manual 1179.0116.02 ─ 02
[SENSe:]FILTer<n>:LPASs:FREQuency[:ABSolute]
<Frequency>
This command selects the absolute low pass filter type for the specified evaluation
For details on the low pass filter refer to
Suffix:
<n>
.
Parameters:
<Frequency>
3kHz | 15kHz | 150kHz
*RST:
15kHz
Default unit: HZ
Example:
FILT:LPAS:FREQ 150kHz
Selects the low pass filter for the demodulation bandwidth range
from 400 kHz to 16 MHz.
Manual operation:
See
[SENSe:]FILTer<n>:LPASs:FREQuency:MANual
<Frequency>
This command selects the cutoff frequency of the low pass filter for the specified evalu-
ation.
For details on the low pass filter refer to
Suffix:
<n>
.
Parameters:
<Frequency>
numeric value
Range:
0 to 3 MHz
*RST:
15kHz
Default unit: HZ
Example:
FILT:LPAS:FREQ:MAN 150kHz
The AF results are restricted to frequencies lower than 150 kHz.
Manual operation:
See
[SENSe:]FILTer<n>:LPASs:FREQuency:RELative
<Frequency>
This command selects the relative low pass filter type for the specified evaluation
For details on the low pass filter refer to
Suffix:
<n>
.
Parameters:
<Frequency>
5PCT | 10PCT | 25PCT
*RST:
25PCT
Default unit: PCT
Configuring the Measurement