General Instrument Setup
R&S
®
FSMR3000
274
User Manual 1179.0116.02 ─ 02
To measure with the accuracy specified in the data sheet, always perform a self-
alignment
when the instrument calls for it.
Starting Touch Screen Alignment
Starts the touchscreen alignment.
Tap the 4 markers on the screen as you are asked to do. The touchscreen is aligned
according to the executed pointing operations.
Alignment Results:
Information on whether the alignment was performed successfully and on the applied
correction data is displayed. The results are available until the next self-alignment
process is started or the instrument is switched off.
Remote command:
10.1.3
How to Perform a Self-Test
You do not have to repeat the self-test every time you switch on the instrument. It is
only necessary when instrument malfunction is suspected.
Operating temperature
Before performing this functional test, make sure that the instrument has reached its
operating temperature (for details, refer to the data sheet).
1. Select [SETUP].
2. Select "Service".
3. Select "Selftest".
Once the instrument modules have been checked successfully, a message is dis-
played.
10.1.4
How to Align the Instrument
Operating temperature
Before performing this functional test, make sure that the instrument has reached its
operating temperature (for details, refer to the data sheet).
To perform a self-alignment
Make sure no signal is connected to the RF input connector. Running a self-alignment
with a signal connected to the RF input can lead to false measurement results.
1. Select [SETUP].
2. Select "Alignment".
Alignment