QuickSave 1 / ... / QuickSave 10
Selects one of the save sets to store the current settings in or to be recalled. At the
time of storage, the "QuickSave 1 / ... / QuickSave 10" placeholder is replaced by a
label indicating the storage date and time and the storage type.
Right-click on one of the QuickSave buttons to display a context menu with additional
functions for the save set.
During recall, save sets of type "Instrument" replace the settings of the entire instru-
ment. All other save sets start a new channel with the stored settings.
Note:
Saving instrument settings in secure user mode.
Settings that are saved via Quick Save in secure user mode are only available during
the current session. As soon as the power is switched off on the R&S
FSMR3, the data
is cleared (see
Chapter 9.2, "Protecting Data Using the Secure User Mode"
Rename ← QuickSave 1 / ... / QuickSave 10
Displays an input field to rename the save set, if write protection is disabled.
Write Protection ← QuickSave 1 / ... / QuickSave 10
Enables or disables write protection for the save set. If enabled, the save set cannot be
renamed or overwritten.
Storage Type (Save only)
Defines which type of settings are stored in the save set.
"Instrument
with all Chan-
nels"
The instrument settings for all currently active "Channel"s are stored.
"Current Chan-
nel"
Only the instrument settings for the currently selected measurement
"Channel"s are stored.
Recall
Restores the instrument settings as saved in the selected settings file. If the settings
file contains settings for a specific "Channel" only, a new channel with the stored set-
tings is activated, otherwise all "Channel"s and instrument settings are overwritten with
the stored settings.
Storing and Recalling Instrument Settings and Measurement Data