Common Analysis and Display Functions
R&S
®
FSMR3000
225
User Manual 1179.0116.02 ─ 02
Remote command:
CALCulate<n>:SPECtrogram:LAYout
Select Frame
Selects a specific frame, loads the corresponding trace from the memory, and displays
it in the Spectrum window.
Note that activating a marker or changing the position of the active marker automati-
cally selects the frame that belongs to that marker.
This function is only available in single sweep mode or if the sweep is stopped, and
only if a spectrogram is selected.
The most recent frame is number 0, all previous frames have a negative number.
Remote command:
CALCulate<n>:SPECtrogram:FRAMe:SELect
History Depth
Sets the number of frames that the R&S
FSMR3 stores in its memory.
If the memory is full, the R&S
FSMR3 deletes the oldest frames stored in the memory
and replaces them with the new data.
Remote command:
CALCulate<n>:SPECtrogram:HDEPth
Time Stamp
Activates and deactivates the timestamp. The timestamp shows the system time while
the measurement is running. In single sweep mode or if the measurement is stopped,
the timestamp shows the time and date of the end of the measurement.
When active, the timestamp replaces the display of the frame number.
Remote command:
CALCulate<n>:SPECtrogram:TSTamp[:STATe]
CALCulate<n>:SPECtrogram:TSTamp:DATA?
Color Mapping
Opens the "Color Mapping" dialog.
Continuous Sweep / Run Cont
While the measurement is running, the "Continuous Sweep" softkey and the [RUN
CONT] key are highlighted. The running measurement can be aborted by selecting the
highlighted softkey or key again. The results are not deleted until a new measurement
is started.
Note:
Sequencer. Furthermore, the [RUN CONT] key controls the Sequencer, not indi-
vidual sweeps. [RUN CONT] starts the Sequencer in continuous mode.
Remote command:
Trace Configuration