Common Analysis and Display Functions
R&S
®
FSMR3000
186
User Manual 1179.0116.02 ─ 02
Unit
The unit in which the y-values of the data points of the limit line are defined.
Compatibility
Indicates whether the limit line definition is compatible with the current measurement
settings.
Visibility
Displays or hides the limit line in the diagram. Up to 8 limit lines can be visible at the
same time. Inactive limit lines can also be displayed in the diagram.
Remote command:
CALCulate<n>:LIMit<li>:LOWer:STATe
CALCulate<n>:LIMit<li>:UPPer:STATe
CALCulate<n>:LIMit<li>:ACTive?
Traces to be Checked
Defines which traces are automatically checked for conformance with the limit lines. As
soon as a trace to be checked is defined, the assigned limit line is active. One limit line
can be activated for several traces simultaneously. If any of the "Traces to be
Checked" violate any of the active limit lines, a message is indicated in the diagram.
Remote command:
CALCulate<n>:LIMit<li>:TRACe<t>:CHECk
Comment
An optional description of the limit line.
X-Offset
Shifts a limit line that has been specified for relative frequencies or times (x-axis) hori-
zontally.
This setting does not have any effect on limit lines that are defined by absolute values
for the x-axis.
Remote command:
CALCulate<n>:LIMit<li>:CONTrol:OFFSet
Y-Offset
Shifts a limit line that has relative values for the y-axis (levels or linear units such as
volt) vertically.
This setting does not have any effect on limit lines that are defined by absolute values
for the y-axis.
Remote command:
CALCulate<n>:LIMit<li>:LOWer:OFFSet
CALCulate<n>:LIMit<li>:UPPer:OFFSet
Create New Line
Creates a new limit line.
Edit Line
Edit an existing limit line configuration.
Result Display Configuration