Common Measurement Settings
R&S
®
FSMR3000
127
User Manual 1179.0116.02 ─ 02
ration. (Note that the frequency is defined as the unsigned value of the equation, thus a
negative frequency is not possible.)
Example: Example for reverse sweep
F
AnalyzerStart
= 100 MHz
F
AnalyzerStop
= 200 MHz
F
Offset
= -300 MHz
Numerator = Denominator = 1
→
F
GeneratorStart
= 200 MHz
→
F
GeneratorStop
= 100 MHz
If the offset is adjusted so that the sweep of the generator crosses the minimum gener-
ator frequency, a message is displayed in the status bar ("Reverse Sweep via min. Ext.
Generator Frequency!").
Example: Example for reverse sweep via minimum frequency
F
AnalyzerStart
= 100 MHz
F
AnalyzerStop
= 200 MHz
F
Offset
= -150 MHz
F
min
= 20 MHz
Numerator = Denominator = 1
→
F
GeneratorStart
= 50 MHz
→
F
GeneratorStop
= 50 MHz via F
min
Displayed Information and Errors
Channel bar
If external generator control is active, some additional information is displayed in the
channel bar.
Label
Description
EXT TG: <source power>
External generator active; signal sent with <source power> level
LVL
Power Offset (see
FRQ
Frequency Offset (see
"(Automatic) Source Frequency (Numerator/Denomi-
Error and status messages
The following status and error messages may occur during external generator control.
Input and Frontend