Spectrum Analyzer
R&S FS300
E-1007.9900.15
6-122
Operating manual, 02/2006
6.2.4.1
Setting the Resolution Bandwidth
Use
The span and the resolution bandwidth can be set independently. In the de-
fault setting, however, these two parameters are coupled to make operating
the instrument easier. This setting is recommended for many applications
because operating errors are prevented. To make certain measurements, it
may be necessary to set the resolution bandwidth and the span independ-
ently. With the R&S FS300, you can manually set resolution bandwidths from
200 Hz to 1 MHz in a 1, 2, 3, 5, 10 sequence.
Setting the
resolution
bandwidth
manually
1.
Press the
function key
in the
menu
.
A selection field containing the available settings is displayed. The de-
fault setting is 1 MHz.
2.
Select a resolution bandwidth with
rotary knob
3.
Press the
ENTER key
[5] to close the selection field.
The new setting is saved and the resolution filter is configured.
Activating automatic
resolution
bandwidth setting
•
Press the
function key
in the
menu
.
The resolution bandwidth is set to
1
/
50
of the span. If the span is changed,
the resolution bandwidth is adjusted automatically.
NOTE
The
function key is not available in the time-domain display mode
(ZERO SPAN).