Operating Manual 1302.6163.12 - 03
6.89
R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
6.5.6
CALCulate:PEAKsearch
I
PSEarch Subsystem
The CALCulate:PEAKsearch subsystem contains commands for verifying scan results in the test
receiver mode.
CALCulate<1|2>:PEAKsearch|PSEarch:AUTO
ON | OFF
By using this command, the peak list in the spurious measurement is calculated automatically after a
measurement. For each range, exactly one peak value is calculated.
The SENSe suffix is unused.
Example:
"CALC:PEAK:AUTO ON"
’Switches the automatic peak search on.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
CALCulate<1|2>:PEAKsearch|PSEarch:ADD
<nummeric value>
This command adds a frequency value to the peaklist in receiver mode (RECEIVER – FINAL MEAS –
EDIT PEAK LIST). Only frequencies in the currently displayed frequency range of the receiver scan
are allowed.
Example:
"CALC:PEAK:ADD 93MHz"
’The frequency 93 MHz as added to the peak list.
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
R
CALCulate<1|2>:PEAKsearch|PSEarch:CLEar[:IMMediate]
The peaklist in receiver mode (RECEIVER – FINAL MEAS – EDIT PEAK LIST) is cleared.
Example:
"CALC:PEAK:CLE"
’The peak list is cleared.
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
R
CALCulate<1|2>:PEAKsearch|PSEarch[:IMMediate]
This command activates the generation of a peak list.
The numeric suffix in
CALCULATE<1|2>
is not significant.
Example:
"CALC:PEAK"
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A, R
This command is an event and therefore has no *RST value and no query.
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...