Operating Manual 1302.6163.12 - 03
4.319
R&S ESU
Instrument Functions
Tracking Generator – Option R&S FSU-B9
4.10 Tracking Generator – Option R&S FSU-B9
During normal operation (without a frequency offset), the tracking generator emits a
signal exactly at the input frequency of the R&S ESU.
For frequency-converting measurements it is possible to set a constant frequency
offset of ±200 MHz between the receive frequency of the R&S ESU and the output
signal of the tracking generator.
Moreover, an I/Q modulation or AM and FM modulation of the output signal can be
provided using two analog input signals.
The output power is level-controlled and can be set in 0.1 dB steps in a range from -
30 dBm to +5 dBm (-100 to + 5 dBm with option R&S FSU-B12).
The tracking generator can be used in all operating modes. Acquisition of test setup
calibration values (SOURCE CAL) and normalization using these correction values
(
NORMALIZE
) is only possible in the
NETWORK
operating mode.
The tracking generator is activated by means of the
NETWORK
hotkey in the hotkey
bar at the bottom of the screen (for details refer to section
.)
4.10.1
Tracking Generator Settings
The
NETWORK
hotkey opens a menu for selecting the functions of the tracking
generator.
NETWORK
The RF characteristics of some DUTs is especially sensitive concerning the input
VSWR. In such cases insertion of 20 dB attenuation between the DUT and the
tracking generator output is highly recommended.
!
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...