General Service Functions
INPUT RF
The
INPUT RF
softkey switches the input of the R&S ESU to the input connector
(normal position).
After
PRESET
,
RECALL
or R&S ESU power on, the
INPUT RF
is always selected.
Remote command:
DIAG:SERV:INP RF
INPUT CAL
The
INPUT CAL
softkey switches the RF input of the R&S ESU to the internal cali-
bration source (128 MHz) and activates the data entry of the output level of the cali-
bration source. Possible values are 0 dB and –30 dB.
Remote command:
DIAG:SERV:INP CAL;
DIAG:SERV:INP:CSO 0 DBM
ENTER
PASSWORD
The
ENTER PASSWORD
softkey allows the entry of a password.
The R&S ESU contains a variety of service functions which, if incorrectly used, can
affect correct operation of the R&S ESU. These functions are normally not accessi-
ble and are only usable after the entry of a password (see instrument service man-
ual).
Remote command:
SYST:PASS "Password"
CAL GEN 128
MHZ
The
CAL GEN 128 MHZ
softkey selects a sinusoidal signal at 128 MHz as output
signal for the internal calibration source. The internal pulse generator will be
switched off.
Remote command:
DIAG:SERV:INP CAL
DIAG:SERV:INP:PULS OFF
CAL GEN
COMB PULSE
This softkey switches the internal pulse generator on and allows the pulse frequency
to be entered.
Available pulse frequencies are 10 kHz, 62.5 kHz, 100 kHz, 1 MHz, 128 MHz, and
640 MHz.
Remote command:
DIAG:SERV:INP:PULS ON;
DIAG:SERV:INP:PULS:PRAT <value>
COMMAND
TRACKING
The
COMMAND TRACKING
softkey activates or deactivates the SCPI error log
function. All remote control commands received by the R&S ESU are recorded in
the following
log file:
D:\R_S\instr\log\IEC_CMDS.LOG
Logging the commands may be extremely useful for debug purposes, e.g. in order
to find misspelled keywords in control programs.
Remote command:
--
If this softkey is activated, the R&S ESU becomes slower.
Summary of Contents for ESU Series
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...