Remote Control
R&S
®
ESR
912
User Manual 1175.7068.02 ─ 12
Suffix:
<n>
.
irrelevant
<generator>
1 | 2
tracking generator
For Noise Figure measurements (K30): 1 only
Parameters:
<Source>
INTernal
the internal reference is used
EXTernal
the external reference is used; if none is available, an error flag
is displayed in the status bar
EAUTo
the external reference is used as long as it is available, then the
instrument switches to the internal reference (Spectrum mode
only)
*RST:
INT
Example:
SOUR:EXT:ROSC EXT
Switches to external reference oscillator
Mode:
A, NF
Manual operation:
See
SOURce<n>:EXTernal<generator>[:STATe]
<State>
This command activates or deactivates the external generator indicated by the suffix
<generator>.
This command is only valid for the External Tracking Generator option (R&S
FSV-B10).
Suffix:
<n>
.
irrelevant
<generator>
1 | 2
external generator
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
SYST:COMM:RDEV:GEN1:TYPE 'SMP02'
Selects SMP02 as generator 1.
"SYST:COMM:RDEV:GEN1:LINK TTL"
Selects GPIB + TTL link as interface.
"SYST:COMM:RDEV:GEN1:ADDR 28"
Sets the generator address to 28.
"SOUR:EXT1:FREQ:SWE ON"
Activates the frequency sweep for generator 1.
"SOUR:EXT ON"
Activates the external generator
Remote Commands to Control the Tracking Generator