Remote Control
R&S
®
ESR
764
User Manual 1175.7068.02 ─ 12
[SENSe:]ESPectrum:RANGe<range>:COUNt
This command returns the number of defined ranges.
Suffix:
<range>
.
1...20
range
Example:
ESP:RANG:COUNt?
Returns the number of defined ranges.
[SENSe:]ESPectrum:RANGe<range>:DELete
This command deletes the specified range. The range numbers are updated accord-
ingly. The reference range cannot be deleted. A minimum of three ranges is manda-
tory.
Suffix:
<range>
.
1...20
range
Example:
ESP:RANG4:DEL
Deletes range 4.
Manual operation:
See
[SENSe:]ESPectrum:RANGe<range>:FILTer:TYPE
<Type>
This command sets the filter type for the specified range.
Note that this parameter can not be set for all ranges if "Fast SEM" mode is activated
(see
Suffix:
<range>
.
1...20
range
Parameters:
<Type>
NORMal
Gaussian filters
CFILter
channel filters
RRC
RRC filters
P5
5 Pole filters
*RST:
NORM
The available bandwidths of the filters are specified in the data
sheet.
Example:
ESP:RANG1:FILT:TYPE RRC
Sets the RRC filter type for range 1.
Remote Commands in Spectrum Analyzer Mode