Remote Control
R&S
®
ESR
763
User Manual 1175.7068.02 ─ 12
●
[SENSe:]ESPectrum:RANGe<range>:INPut:
●
Preamp, see
[SENSe:]ESPectrum:RANGe<range>:INPut:GAIN:STATe
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
ESP:HSP ON
Manual operation:
See
[SENSe:]ESPectrum:RANGe<range>:BANDwidth[:RESolution]
<Value>
This command sets the RBW value for the specified range.
Note that this parameter can not be set for all ranges if "Fast SEM" mode is activated
(see
Suffix:
<range>
.
1...20
range
Parameters:
<Value>
Refer to the data sheet.
*RST:
30.0 kHz
Example:
ESP:RANG2:BAND:RES 5000
Sets the RBW for range 2 to 5 kHz.
Manual operation:
See
[SENSe:]ESPectrum:RANGe<range>:BANDwidth:VIDeo
<Value>
This command sets the VBW value for the specified range.
Note that this parameter can not be set for all ranges if "Fast SEM" mode is activated
(see
Suffix:
<range>
.
1...20
range
Parameters:
<Value>
Refer to the data sheet.
*RST:
10.0 MHz
Example:
ESP:RANG1:BAND:VID 5000000
Sets the VBW for range 1 to 5 MHz.
Manual operation:
See
Remote Commands in Spectrum Analyzer Mode