Remote Control
R&S
®
ESR
683
User Manual 1175.7068.02 ─ 12
[SENSe:][WINDow:]DETector<trace>[:FUNCtion]
<Detector>
This command selects the detector for the scan.
Suffix:
<trace>
.
1...6
Selects the trace.
Parameters:
<Detector>
NEGative | POSitive | RMS | AVERage | QPEak | CAVerage |
CRMS
*RST:
APEak
Example:
DET RMS
Selects the RMS detector.
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:MODE
<Mode>
This command defines the type of display and the evaluation of the traces. WRITE cor-
responds to the Clr/Write mode of manual operation. The trace is switched off (=
BLANK in manual operation) with
DISPlay[:WINDow<n>]:TRACe<t>[:STATe]
.
The number of measurements for AVERage, MAXHold and MINHold is defined with
the
or
be noted that synchronization to the end of the indicated number of measurements is
only possible in single sweep mode.
If calculation of average values is active, selection between logarithmic and linear aver-
aging is possible. For more detail see
Suffix:
<n>
.
window; For applications that do not have more than 1 measure-
ment window, the suffix <n> is irrelevant.
<t>
trace
Parameters:
<Mode>
For more information on trace modes see
AVERage
Average trace mode.
BLANk
Blank trace.
MAXHold
Maximum value trace.
MINHold
Minimum value trace.
VIEW
Frozen trace.
Remote Commands in Receiver Mode