Quick Start
R&S
®
ESR
60
User Manual 1175.7068.02 ─ 12
3. Press the "GPIB" softkey.
The submenu for setting the parameters of the remote control interface is dis-
played.
Setting the GPIB address
► In the "GPIB" menu, press the "GPIB Address" softkey.
The edit dialog box for the GPIB address is displayed. The setting range is from 0
to 30. If SCPI is selected as the GPIB language, the default address is 20.
Remote command:
SYST:COMM:GPIB:ADDR 20
Setting the default ID response string
► In the "GPIB" menu, press the "ID String Factory" softkey to select the default
response to the
*IDN?
command.
Setting a user-defined ID response string
► In the "GPIB" menu, press the "ID String User" softkey to enter a user-defined
response to the
*IDN?
command. The maximum length of the output string is 36
characters.
3.3
Firmware Update and Installation of Firmware
Options
This chapter describes how to update the firmware and how to activate optional firm-
ware packages.
Updating the firmware or installing optional firmware requires administrator rights (see
3.3.1
Firmware Update
You can install a new firmware version using one of the following methods:
●
Copying the files to the instrument using USB devices (e.g. a memory stick), GPIB,
or LAN
●
Using the "Remote Installation" feature in a LAN network
Copying the files to the instrument
1. Press the [SETUP] key.
2. Press the "More" softkey.
Firmware Update and Installation of Firmware Options