![R&S CMW-KM5 Series User Manual Download Page 637](http://html1.mh-extra.com/html/rands/cmw-km5-series/cmw-km5-series_user-manual_3148802637.webp)
LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
637
User Manual 1173.9628.02 ─ 16
Parameters:
<FrequencyError>
Range:
0 ppm to 1 ppm
*RST:
0.1 ppm, ON
Default unit: ppm
Additional parameters: OFF | ON (disables the limit check | ena-
bles the limit check using the previous/default limit values)
Example:
See
Firmware/Software:
V1.0.10.1
CONFigure:LTE:MEAS<i>:MEValuation:LIMit:QAM<ModOrder>:IQOFfset
<Enable>, <Offset1>, <Offset2>, <Offset3>
Defines upper limits for the I/Q origin offset (16-QAM or 64-QAM modulation). Three
different I/Q origin offset limits can be set for three TX power ranges. For details see
chapter 3.2.5.4, "I/Q Origin Offset Limits"
Suffix:
<ModOrder>
.
16,64
16-QAM or 64-QAM
Parameters:
<Enable>
OFF | ON
OFF
: disables the limit check
ON
: enables the limit check
*RST:
ON
<Offset1>
I/Q origin offset limit for high TX power range
Range:
-256 dBc to 256 dBc
*RST:
-24.2 dBc
Default unit: dBc
<Offset2>
I/Q origin offset limit for intermediate TX power range
Range:
-256 dBc to 256 dBc
*RST:
-19.2 dBc
Default unit: dBc
<Offset3>
I/Q origin offset limit for low TX power range
Range:
-256 dBc to 256 dBc
*RST:
-9.2 dBc
Default unit: dBc
Example:
See
Firmware/Software:
V1.0.10.1
Command Reference