![R&S CMW-KM5 Series User Manual Download Page 620](http://html1.mh-extra.com/html/rands/cmw-km5-series/cmw-km5-series_user-manual_3148802620.webp)
LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
620
User Manual 1173.9628.02 ─ 16
Example:
See
Firmware/Software:
V2.0.10
V2.1.25: increased maximum number of segments to 250
V2.1.30: increased maximum number of segments to 512
V3.0.50: increased maximum number of segments to 1000
Options:
R&S CMW-KM012
Manual operation:
See
"List Mode > Offline Segment No."
CONFigure:LTE:MEAS<i>:MEValuation:LIST:SEGMent<no>:SETup
<SegmentLength>, <Level>, <DuplexMode>, <Band>, <Frequency>,
<ChBandwidth>, <CyclicPrefix>, <ChannelType>, <RetriggerFlag>,
<EvaluatOffset>[, <NetworkSigValue>]
Defines the length and analyzer settings of segment <no>. In general this command
must be sent for all segments measured (
).
.
Suffix:
<no>
.
1..2000
Segment number
Parameters:
<SegmentLength>
Number of subframes in the segment
Range:
1 to 2000
*RST:
1
<Level>
Expected nominal power in the segment. The range can be cal-
culated as follows:
Range (Expected Nominal Power) = Range (Input Power) +
External Attenuation - User Margin
Range:
-47 dBm to 55 dBm for the input power at the RF
COM connectors (please notice also the ranges
quoted in the data sheet)
*RST:
0 dBm
Default unit: dBm
<DuplexMode>
FDD | TDD
Duplex mode used in the segment
<Band>
FDD: OB1 | ... | OB28 | OB30 | OB31
TDD: OB33 | ... | OB44
Operating Band (OB) used in the segment
*RST:
OB1 (OB33 for TDD)