LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
615
User Manual 1173.9628.02 ─ 16
CONFigure:LTE:MEAS<i>:MEValuation:SCOunt:SPECtrum:ACLR
<StatisticCount>
CONFigure:LTE:MEAS<i>:MEValuation:SCOunt:SPECtrum:SEMask
<StatisticCount>
Specifies the statistic count of the measurement. The statistic count is equal to the
number of measurement intervals per single shot. Use
CONFigure:...:MEAS<i>:...:REPetition SINGleshot | CONTinuous
to
select either single-shot or continuous measurements.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "Remote Con-
trol"
Separate statistic counts for ACLR and spectrum emission mask measurements are
supported.
Parameters:
<StatisticCount>
Number of measurement intervals (slots)
Range:
1 slot to 1000 slots
*RST:
20 slots
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V1.0.10.1
Manual operation:
See
"Emission Mask / ACLR > Statistic Count"
CONFigure:LTE:MEAS<i>:MEValuation:SPECtrum:SEMask:MFILter
<MeasFilter>
Selects the resolution filter type for filter bandwidths of 100 kHz and 1 MHz. For 30 kHz
filters the type is fixed (gaussian shape).
Parameters:
<MeasFilter>
BANDpass | GAUSs
*RST:
BAND
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V1.0.15.20
Manual operation:
See
CONFigure:LTE:MEAS<i>:MEValuation:SPECtrum:ACLR:ENABle
<UTRA1>,
<UTRA2>, <EUTRA>
Enables or disables the evaluation of 1
st
adjacent UTRA channels, 2
nd
adjacent UTRA
channels and 1
st
adjacent E-UTRA channels.
Parameters:
<UTRA1>
OFF | ON
*RST:
ON
<UTRA2>
OFF | ON
*RST:
ON
Command Reference