LTE Multi Evaluation Measurement
R&S
®
CMW-KM5xx/-KS5xx
603
User Manual 1173.9628.02 ─ 16
CONFigure:LTE:MEAS<i>:MEValuation:TMODe:ENPower
<ExpNomPow>...
Defines the expected nominal power values for all entries of the "TPC Mode" list.
For definition of the corresponding subframe count values, see
.
Parameters:
<ExpNomPow>
Comma separated list of 16 values, for list entry number 0 to 15
The range of the expected nominal power can be calculated as
follows:
Range (Expected Nominal Power) = Range (Input Power) +
External Attenuation - User Margin
Range:
-47 dBm to 55 dBm for the input power at the RF
COM connectors (please notice also the ranges
quoted in the data sheet)
*RST:
0 dBm
Default unit: dBm
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V2.1.30
Manual operation:
See
CONFigure:LTE:MEAS<i>:MEValuation:TMODe:RLEVel?
Queries the reference level for all entries of the "TPC Mode" list. The reference level is
calculated from the expected nominal power of each entry and the user margin.
Return values:
<ReferenceLevel>
Comma separated list of 16 values, for list entry number 0 to 15
The range of the reference levels can be calculated as follows:
Range (Reference Level) = Range (Input Power) + External
Attenuation
Range:
-47 dBm to 55 dBm for the input power at the RF
COM connectors (please notice also the ranges
quoted in the data sheet)
Default unit: dBm
Example:
See
Specifying Additional Measurement-Specific Settings
Usage:
Query only
Firmware/Software:
V2.1.30
Manual operation:
See
CONFigure:LTE:MEAS<i>:MEValuation:MOEXception
<MeasOnException>
Specifies whether measurement results that the R&S
CMW identifies as faulty or inac-
curate are rejected.