When the measurement has completed the first measurement cycle (first single shot),
the statistical depth is reached and the timer is reset.
If the first measurement cycle has not been completed when the timer expires, the
measurement is stopped. The measurement state changes to
RDY
and the reliability
indicator is set to 1, indicating that a measurement timeout occurred. Still running
READ
,
FETCh
or
CALCulate
commands are completed, returning the available results.
At least for some results there are no values at all or the statistical depth has not been
reached.
A timeout of 0 s corresponds to an infinite measurement timeout.
Parameters:
<Timeout>
Default unit: s
Firmware/Software:
V2.0.10
CONFigure:LTE:SIGN<i>:EBLer:REPetition
<Repetition>
Specifies whether the measurement is stopped after a single shot or repeated continu-
ously.
Parameters:
<Repetition>
SINGleshot | CONTinuous
SINGleshot
: Single-shot measurement
CONTinuous
: Continuous measurement
*RST:
SING
Example:
See
Configuring a BLER Measurement
Firmware/Software:
V3.0.30
Manual operation:
See
CONFigure:LTE:SIGN<i>:EBLer:SCONdition
<StopCondition>
Selects whether a BLER measurement without stop condition or a confidence BLER
measurement with early decision concept is performed.
Parameters:
<StopCondition>
NONE | CLEVel
NONE
: no stop condition, no early termination of measurement
CLEVel
: confidence BLER measurement
*RST:
NONE
Example:
See
Configuring a BLER Measurement
Firmware/Software:
V3.0.30
Options:
R&S CMW-KS510 for CLEVel
Manual operation:
See
Command Reference