LTE Signaling
R&S
®
CMW-KM5xx/-KS5xx
422
User Manual 1173.9628.02 ─ 16
Example:
See
Firmware/Software:
V1.0.15.20
V2.0.20: additional value ZERO
V2.1.20: additional values N17 and T12
V3.0.20: additional values N9, N54 and T18
V3.2.70: additional values N32, N45 and N60
Manual operation:
See
CONFigure:LTE:SIGN<i>:CONNection[:PCC]:RMC:RBPosition:DL<Stream>
<Position>
CONFigure:LTE:SIGN<i>:CONNection:SCC:RMC:RBPosition:DL<Stream>
<Position>
Select the position of the allocated resource blocks within the channel bandwidth for
PCC or SCC. The command configures one downlink stream. Set the same value for
both streams of a carrier.
The RBs can always be located at the lower end, starting with RB number 0, or at the
upper end of the channel. The other values are only allowed for certain one TX
antenna configurations, see
chapter 2.2.12.2, "DL RMCs, One TX Antenna"
Suffix:
<Stream>
.
1..2
Selects the downlink stream. Can be omitted for transmission
schemes with only one downlink stream per carrier.
Parameters:
<Position>
LOW | HIGH | P5 | P10 | P23 | P35 | P48
*RST:
LOW
Example:
See
Firmware/Software:
V1.0.15.21
V2.0.10: <Stream> added
V2.1.20: new values P5 | P10 | P23 | P35 | P48
V3.2.50: added SCC command
Options:
R&S CMW-KS502/-KS552 (FDD/TDD) for SCC
Manual operation:
See
CONFigure:LTE:SIGN<i>:CONNection[:PCC]:RMC:RBPosition:UL
<Position>
Select the position of the allocated resource blocks within the channel bandwidth for
uplink.
The RBs can always be located at the lower end, starting with RB number 0, or at the
upper end of the channel.
The other values are only allowed for certain RMC configurations, see
Command Reference