LTE Signaling
R&S
®
CMW-KM5xx/-KS5xx
238
User Manual 1173.9628.02 ─ 16
// *****************************************************************************
CONFigure:LTE:SIGN:CONNection:PCC:RMC:DL N50,QPSK,T5
CONFigure:LTE:SIGN:CONNection:PCC:RMC:UL N50,QPSK,T6
CONFigure:LTE:SIGN:CONNection:PCC:RMC:RBPosition:DL LOW
CONFigure:LTE:SIGN:CONNection:PCC:RMC:RBPosition:UL LOW
// *****************************************************************************
// Modify the RS EPRE level for the PCC.
// *****************************************************************************
CONFigure:LTE:SIGN:DL:PCC:RSEPre:LEVel -83
// *****************************************************************************
// Command the UE to change the UL power by +3 dB:
// Select single pattern as active TPC setup, define a single pattern of
// 3 steps UP, execute the pattern.
// *****************************************************************************
CONFigure:LTE:SIGN:UL:PUSCh:TPC:SET SINGle
CONFigure:LTE:SIGN:UL:PUSCh:TPC:SINGle 3, UP
CONFigure:LTE:SIGN:UL:PUSCh:TPC:PEXecute
// *****************************************************************************
// Apply a continuous TPC command pattern of 5x -1 dB and 4x +1 dB:
// Define the pattern and select the active TPC setup.
// *****************************************************************************
CONFigure:LTE:SIGN:UL:PUSCh:TPC:UDPattern 9,-1,-1,-1,-1,-1,1,1,1,1
CONFigure:LTE:SIGN:UL:PUSCh:TPC:SET UDContinuous
// *****************************************************************************
// Command the UE to a target power of -10 dBm.
// *****************************************************************************
CONFigure:LTE:SIGN:UL:PUSCh:TPC:CLTPower -10
CONFigure:LTE:SIGN:UL:PUSCh:TPC:SET CLOop
// *****************************************************************************
// Execute ramping up pattern B for a 3GPP relative power control test.
// *****************************************************************************
CONFigure:LTE:SIGN:UL:PUSCh:TPC:SET RPControl
CONFigure:LTE:SIGN:UL:PUSCh:TPC:RPControl RUB
CONFigure:LTE:SIGN:UL:PUSCh:TPC:PEXecute
2.5.2 BLER Tests
The BLER measurement provided by the LTE signaling application is programmed as
follows:
●
The application is controlled by SCPI commands with the following syn-
tax:
...:LTE:SIGN:EBLer:...
●
After a
*RST
, the measurement is switched off.
Programming
深圳德标仪器