CDMA2000 Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
807
User Manual 1173.9511.02 ─ 11
Example:
See
Multi Evaluation: Specifying Limits
Firmware/Software:
V1.0.2.7
Manual operation:
See
CONFigure:CDMA:MEAS<i>:MEValuation:LIMit:PERRor
<PErrRMS>, <PErrPeak>
Defines a symmetric limit for the RMS and peak values of the phase error. The limit
check fails if the absolute value of the measured phase error exceeds the specified
value.
Parameters:
<PErrRMS>
Range:
0 deg to 180 deg
*RST:
13.6 deg
Default unit: deg
Additional parameters: OFF | ON (disables the limit check | ena-
bles the limit check using the previous/default limit values)
<PErrPeak>
Range:
0 deg to 180 deg
*RST:
19.6 deg
Default unit: deg
Additional parameters: OFF | ON (disables the limit check | ena-
bles the limit check using the previous/default limit values)
Example:
See
Multi Evaluation: Specifying Limits
Firmware/Software:
V1.0.2.7
Manual operation:
See
CONFigure:CDMA:MEAS<i>:MEValuation:LIMit:IQOFfset
<IQOffset>
Defines an upper limit for the I/Q origin offset (Carrier Feedthrough).
Parameters:
<IQOffset>
Range:
-120 dB to -20 dB
*RST:
-25 dB
Default unit: dB
Additional parameters: OFF | ON (disables the limit check | ena-
bles the limit check using the previous/default limit values)
Example:
See
Multi Evaluation: Specifying Limits
Firmware/Software:
V1.0.2.7
Manual operation:
See
CONFigure:CDMA:MEAS<i>:MEValuation:LIMit:IQIMbalance
<IQImbalance>
Defines an upper limit for the I/Q imbalance.