CDMA2000 Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
797
User Manual 1173.9511.02 ─ 11
CONFigure:CDMA:MEAS<i>:MEValuation:SCOunt:SPECtrum
<SCountSpectrum>
Specifies the statistic count of the measurement. The statistic count is equal to the
number of measurement intervals per single shot. Use
CONFigure:...:MEAS<i>:...:REPetition SINGleshot | CONTinuous
to
select either single-shot or continuous measurements.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "Remote Con-
trol"
Parameters:
<SCountSpectrum>
Number of measurement intervals.
Range:
1 to 1000
*RST:
1
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Firmware/Software:
V1.0.2.7
V2.0.10: Change *RST value to 1.
Manual operation:
See
"Statistic Count Modulation / Spectrum"
CONFigure:CDMA:MEAS<i>:MEValuation:MOEXception
<MeasOnException>
Specifies whether measurement results that the R&S
CMW identifies as faulty or inac-
curate are rejected.
Parameters:
<MeasOnException> OFF | ON
ON:
Results are never rejected
OFF:
Faulty results are rejected
*RST:
OFF
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Firmware/Software:
V1.0.2.7
Manual operation:
See
CONFigure:CDMA:MEAS<i>:MEValuation:SFACtor
<SpreadingFactor>
Selects the spreading factor for the code domain power and code domain error meas-
urements.
Parameters:
<SpreadingFactor>
SF16 | SF32 | SF64
SF16:
spreading factor 16
SF32:
spreading factor 32
SF64:
spreading factor 64
*RST:
SF32
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Firmware/Software:
V1.0.2.7
Command Reference
深圳德标仪器