CDMA2000 Signaling
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
607
User Manual 1173.9511.02 ─ 11
CONFigure:CDMA:SIGN<i>:MSINfo:MEID?
......................................................................608
CONFigure:CDMA:SIGN<i>:TEST:MSINfo:MEID
.............................................................. 609
CONFigure:CDMA:SIGN<i>:MSINfo:MSUPport?
...............................................................609
CONFigure:CDMA:SIGN<i>:MSINfo:PREVision?
.............................................................. 609
SENSe:CDMA:SIGN<i>:ATADdress:IPV<n>?
...................................................................610
SENSe:CDMA:SIGN<i>:TEST:RX:POWer:STATe?
Queries the the quality of the RX signal from the connected MS.
Return values:
<State>
NAV | LOW | OK | HIGH
NAV
: no signal from MS detected
LOW
: the MS power is below the expected range
OK
: the MS power is in the expected range
HIGH
: the MS power is above the expected range
Example:
See
Usage:
Query only
Firmware/Software:
V3.2.70
Manual operation:
See
CONFigure:CDMA:SIGN<i>:MSINfo:DNUMber?
Queries the number dialed at the MS.
Return values:
<DialedNumber>
Dialed number as string.
Example:
See
Usage:
Query only
Firmware/Software:
V1.0.15.0
Manual operation:
See
CONFigure:CDMA:SIGN<i>:MSINfo:EIRP?
Queries the information from the MS about the Maximum Effective Isotropic Radiated
Power (EIRP).
Return values:
<MaxEIRP>
Range:
0 to 999
Example:
See
Usage:
Query only
Firmware/Software:
V1.0.15.0
Manual operation:
See
Command Reference