1xEV-DO Signaling
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
220
User Manual 1173.9511.02 ─ 11
INITiate:EVDO:SIGN<i>:THRoughput
STOP:EVDO:SIGN<i>:THRoughput
ABORt:EVDO:SIGN<i>:THRoughput
These remote commands control the Throughput (Performance) measurement - either
forward or reverse, depending on the current Test Application mode (see
).
Starts, stops, or aborts the measurement:
●
INITiate...
starts or restarts the measurement; the R&S
CMW enters the
"RUN" state.
●
STOP...
causes a running measurement to stop after the current evaluation
period is terminated and valid results are available; the R&S
CMW enters the
"RDY" state.
●
ABORt...
causes a running measurement to stop immediately; the R&S
CMW
enters the "OFF" state.
Use
FETCh...STATe?
to query the current measurement state.
See also: "Measurement Control" in the R&S
CMW user manual, chapter "Remote
Control"
Example:
See
Perform Forward Link Measurements
Usage:
Event
Firmware/Software:
V2.1.20
Options:
R&S CMW-KS890
Manual operation:
See
CONFigure:EVDO:SIGN<i>:RXQuality:PER:TOUT
<Timeout>
CONFigure:EVDO:SIGN<i>:RXQuality:THRoughput:TOUT
<Timeout>
Defines a timeout for the measurement. The timer is started when the measurement is
initiated via a
READ
or
INIT
command. It is not started if the measurement is initiated
manually (ON/OFF key or RESTART/STOP key).
When the measurement has completed the first measurement cycle (first single shot),
the statistical depth is reached and the timer is reset.
If the first measurement cycle has not been completed when the timer expires, the
measurement is stopped. The measurement state changes to
RDY
and the reliability
indicator is set to 1, indicating that a measurement timeout occurred. Still running
READ
,
FETCh
or
CALCulate
commands are completed, returning the available results.
At least for some results there are no values at all or the statistical depth has not been
reached.
A timeout of 0 s corresponds to an infinite measurement timeout.
Parameters:
<Timeout>
Default unit: s
Firmware/Software:
V3.0.10
Command Reference