![R&S AREG800A User Manual Download Page 91](http://html2.mh-extra.com/html/rands/areg800a/areg800a_user-manual_3080572091.webp)
Configuring the measurement setup
R&S
®
AREG800A
91
User Manual 1179.3615.02 ─ 01
Displayed are channel ID and channel name for each radar channel. You can check
the location of the output connectors of the radar channel by clicking "Show Connec-
tor". Also, you can select the mapping frontend and the mapping sensor, adjust levels
of each radar channel individually or adjust all radar channel levels at the same time.
ID ← Channel mapping table
Displays the identification name of the radar channel.
Displayed are eight radar channels that are designated "A1" to "D2". For details, see
Remote command:
[:SOURce<hw>]:AREGenerator:CHANnel:ID?
Alias/Channel ← Channel mapping table
Sets the alias of the radar channel, that is the channel name.
Also, displays this channel name in the "Channel Mapping" table in the "Channel" col-
umn.
Remote command:
[:SOURce<hw>]:AREGenerator:CHANnel:NAME
Show Connector ← Channel mapping table
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
Mapping Frontend ← Channel mapping table
Selects the external frontend and its channel that are mapped to the radar channel.
Remote command:
[:SOURce<hw>]:AREGenerator:MAPPing<ch>:FE
Mapping Sensor ← Channel mapping table
Selects the sensor that is mapped to the radar channel.
Remote command:
[:SOURce<hw>]:AREGenerator:MAPPing<ch>:SENSor
Adjust Level ← Channel mapping table
Adjusts the level of a radar channel automatically.
Remote command:
[:SOURce<hw>]:AREGenerator:MAPPing<ch>:ADJust:LEVel
Adjust All Levels
Adjusts the levels of all radar channels automatically.
Remote command:
[:SOURce<hw>]:AREGenerator:MAPPing<ch>:ADJust:ALL
Adjust Level Settings
Accesses a dialog to adjust level settings. See
Chapter 6.5.2, "Adjust Level settings"
Channel mapping