Network operation and remote control
R&S
®
AREG800A
194
User Manual 1179.3615.02 ─ 01
larly since faulty commands from the controller to the instrument are recorded there as
well.
10.4.8
Reset values of the status reporting system
The following table contains the different commands and events causing the status
reporting system to be reset. None of the commands, except of
*RST
and
SYSTem:PRESet
affect the functional instrument settings. In particular,
DCL
does not
change the instrument settings.
Table 10-7: Resetting the status reporting system
Event
Switching on supply
voltage
Power-On-Status-
Clear
DCL,
SDC
(Device
Clear,
Selected
Device
Clear)
*RST
or
SYSTem:
PRESet
STATus:
PRESet
*CLS
Effect
0
1
Clear STB, ESR
-
Yes
-
-
-
Yes
Clear SRE, ESE
-
Yes
-
-
-
-
Clear PPE
-
Yes
-
-
-
-
Clear error queue
Yes
Yes
-
-
-
Yes
Clear output buffer
Yes
Yes
Yes
1)
1)
1)
Clear command processing
and input buffer
Yes
Yes
Yes
-
-
-
1) The first command in a command line that immediately follows a <PROGRAM MESSAGE TERMINA-
TOR> clears the output buffer.
10.5
Remote access settings
This section outlines the settings required for accessing and configuring the provided
remote control interfaces. It includes network settings, access addresses, emulation
settings for using the command sets of other generators, and the access whith smart
devices.
About instrument emulations
You can remotely control the R&S
AREG800A using the command set of another sig-
nal generator. With this function you can, for example, replace a signal generator with
an R&S
AREG800A in an automated test setup, without adjusting the command scripts
used.
You find the remote control command sets supported by the R&S
AREG800A in a
selection list.
Remote access settings